• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC UNK - NAILS: TFN; ROD, FIXATION, INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC UNK - NAILS: TFN; ROD, FIXATION, INTRAMEDULLARY Back to Search Results
Device Problem Break (1069)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
This report is for an unk - nails: tfn/unknown lot.Part and lot numbers are unknown; udi number is unknown.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Without a lot number, the device history records review could not be completed as no product was received.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2020, an unknown trochanteric fixation nail broke.The tfna revision due to a non-union, possible issue with the nail breaking.The unknown tfn nail was removed yesterday, and a new nail was put in.The procedure successfully completed.The patient outcome is unknown.Concomitant device reported: unk-helical blades (part# unknown; lot# unknown; quantity: 1).Unk-screws (part# unknown; lot# unknown; quantity: unknown).This complaint involves one (1) device.This report is for (1) unk - nails: tfn.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h4, h6: the complaint product is not available to return for investigation.The customer retaining the product.H6: patient code 3191 used to capture additional medical/surgical intervention required.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d11: concomitant product added to complaint.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK - NAILS: TFN
Type of Device
ROD, FIXATION, INTRAMEDULLARY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10634742
MDR Text Key209984912
Report Number2939274-2020-04505
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 09/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received10/09/2020
Patient Sequence Number1
Treatment
UNK - NAIL HEAD ELEMENTS: TFN HELICAL BLADE.; UNK - NAIL HEAD ELEMENTS: TFN HELICAL BLADE.; UNK - NAILS: TFN.; UNK - SCREWS: TRAUMA.; UNK - SCREWS: TRAUMA.; UNK - NAIL HEAD ELEMENTS: TFN HELICAL BLADE; UNK - SCREWS: TRAUMA
Patient Outcome(s) Required Intervention;
Patient Age75 YR
Patient Weight87
-
-