RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Incorrect, Inadequate or Imprecise Resultor Readings (1535); Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
Malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.
Date of this report: 06 oct 2020.
|
|
Event Description
|
It was reported that the ventilator had a battery fault during testing.
There was no patient involvement.
The service engineer (se) inspected the device.
The se found a battery failed alarm and re-connect the battery, but the problem still existed.
The se replaced the battery to address the reported issue.
The unit was checked overall, run in tested, cleaned, functionally tested and no abnormality was confirmed.
|
|