• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG AS ENDURO TIBIA HEMI-WEDGE T3 4MM RL/LM; KNEE ENDOPROSTHETICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG AS ENDURO TIBIA HEMI-WEDGE T3 4MM RL/LM; KNEE ENDOPROSTHETICS Back to Search Results
Model Number NB075Z
Device Problem Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation on going.Additional information / results will be submitted in a supplemental report.
 
Event Description
It was reported that there was an issue with as enduro tibia.According to the description text, it was reported that the inner packing is damaged.There was no patient harm reported.Additional information was not provided nor available / was not available.Additional patient information is not available.The adverse event / malfunction is filed under aag reference (b)(4).Associated medwatch-reports: 9610612-2020-00651 (400486980 nb065z).
 
Manufacturer Narrative
Investigation results: visual investigation: the packaging processes in the responsible production department are validated.The packages itself will be reviewed by 100 per cent visual inspection within the production process.Therefore it is unlikely that the packages have left the production department in such a damaged condition.There are no hints that the packages was not sealed in a proper way.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 4 (5) probability of occurrence) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results there is capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS ENDURO TIBIA HEMI-WEDGE T3 4MM RL/LM
Type of Device
KNEE ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key10642925
MDR Text Key210715050
Report Number9610612-2020-00659
Device Sequence Number1
Product Code KRO
Combination Product (y/n)N
PMA/PMN Number
K101815
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNB075Z
Device Catalogue NumberNB075Z
Device Lot Number52322218
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2020
Date Manufacturer Received12/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-