• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL HEAD Back to Search Results
Catalog Number UNK HIP FEMORAL HEAD
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); No Code Available (3191)
Event Date 09/22/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Left hip revision for infection.Product codes and lot codes are unavailable no further patient details are available and no previous operation record is available.Doe: (b)(6) 2020; left hip.Corail collarless stem.Cathcart spacer.Cathcart head.
 
Manufacturer Narrative
Product complaint # (b)(4).No device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL HEAD
Type of Device
HIP FEMORAL HEAD
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10643111
MDR Text Key210265445
Report Number1818910-2020-21712
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL HEAD
Was Device Available for Evaluation? No
Date Manufacturer Received10/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN HIP FEMORAL AUGMENT; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP FEMORAL STEM; UNKNOWN HIP FEMORAL AUGMENT; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP FEMORAL STEM
Patient Outcome(s) Required Intervention;
Patient Age73 YR
-
-