MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PROMUS ELEMENT PLUS; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PROMUS ELEMENT PLUS; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 9390
Device Problem Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/11/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr.: promus element plus,mr,ous 2.25x28mm stent delivery system was returned for analysis.A visual examination of the stent found proximal stent damage.The undamaged stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon was reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the shaft polymer extrusion found no issues.An examination (visual and via scope) found no issues.0.0140" recommended guidewire inserted through distal tip and exited at exchange port without issue.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 21-sep-2020.It was reported that advancing difficulties were encountered.The target lesion was located in the left anterior descending artery.A 2.25x28mm promus element plus drug eluting stent was advanced for treatment.However, the device could not be advanced to the lesion part.The procedure was completed with another of the same device.There were no patient complications reported and the patient was stable.However, returned device analysis revealed stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS ELEMENT PLUS
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10643645
MDR Text Key210415782
Report Number2134265-2020-13321
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/09/2022
Device Model Number9390
Device Catalogue Number9390
Device Lot Number0025222962
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/21/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient Weight65
-
-