Model Number V60
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 07oct2020.
 
Event Description
It was reported to philips that the device had frozen up.The device was not being used on a patient at the time of the event.There was no patient or user harm reported.
 
Manufacturer Narrative
G4:10dec2020.B4:11dec2020.It was reported to philips that the device had a +35 volt supply failure.The device was evaluated by the customer with assistance from a philips remote service engineer (rse).And it was determined that the motor control printed circuit board assembly (mc pcba) needed to be replaced to return the device to working condition.The customer's bio-med replaced the mc pcba to resolve the issue and bring the device back to functionality.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key10645872
MDR Text Key212484756
Report Number2031642-2020-03624
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 09/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-