RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Failure to Charge (1085); Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 07oct2020.
|
|
Event Description
|
It was reported that the unit alarmed to check the battery.There was no patient involvement.The field service engineer (fse) provided remote support and through troubleshooting determined the backup battery was not charging.The fse instructed the customer to run on battery power per the pvt battery test and if after 20 minutes of running on battery the battery does not charge then to replace the power management board.
|
|
Manufacturer Narrative
|
Following the evaluation of the device, the field service engineer (fse) recommended replacing the battery.Multiple attempts were made to follow up with the customer to obtain repair information; however, there was no response.If additional information is received at a later date, a supplemental report will be submitted.
|
|