Device evaluated by mfr.: synergy ous mr 3.00 x 28mm stent delivery system was returned for analysis.A visual examination of the stent found signs of stent damage.A stent strut in the proximal section of the stent lifted and pulled distally.The undamaged stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no issues.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.A 0.014" guidewire loaded successfully through the distal end of the tip and exited at the exchange port without issue.No other issues were identified during the product analysis.
|