MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
|
Back to Search Results |
|
Model Number MMT-1715K |
Device Problems
Obstruction of Flow (2423); No Apparent Adverse Event (3189)
|
Patient Problems
Bruise/Contusion (1754); Hematoma (1884); Complaint, Ill-Defined (2331); Blood Loss (2597)
|
Event Date 09/29/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that they were hospitalized for unknown reason on unknown date.Blood glucose reading was unknown.It was unknown that how the customer was treated at the hospital.Customer stated that insulin pump had insulin flow block alarm and customer was changed infusion set and reservoir, but issue was not resolved.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Device passed the functional tests, including the self test, sleep current measurement, active current measurement, rewind test, prime or seating test, basic occlusion test, occlusion test, force sensor test, displacement test and delivery accuracy test at 0.08720 inches.No unexpected insulin flow blocked alarm noted.Device received with pillowing keypad overlay and cracked retainer.(b)(4).
|
|
Manufacturer Narrative
|
This report is part of a retrospective review and remediation efforts in response to a warning letter.Updated h9: z-0955-2020.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Search Alerts/Recalls
|
|
|