Model Number 71335558
Device Problem Connection Problem (2900)
Patient Problems Injury (2348); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/17/2020
Event Type  Injury  
Event Description
It was reported that, during surgery, after inserting the acetabulum cup 58mm, a 58mm poly liner compatible with the 58mm cup was tried to be used but could not be inserted.A second liner was tried (58mm 20 degree elevation type), but the same problem occurred.Finally, additional reaming was performed and a 60mm cup was used to complete the procedure.A delay greater than 30 minutes occurred without significant effect on the patient.
 
Event Description
It was reported that during surgery, after inserting the acetabulum cup 58mm, a 58mm poly liner compatible with the 58mm cup was tried to be used but could not be inserted, a second liner was tried (58mm 20 degree elevation type) but would still not work and a third liner was tried but the same problem occurred.Finally additional reaming was performed and a 60mm cup was used to complete the procedure.A delay greater de 30 minutes occurred without significant effect on the patient.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection could not confirm the stated failure.The device exhibits normal wear and use from attempted insertion of the implant.Dimensional evaluation was completed on the returned shell device for all ctq features critical to liner mating and function.No deviations were noted during dimensional evaluation.The failure mode cannot be confirmed via dimensional analysis of the device.A medical investigation was conducted and confirms per complaint the details, we are currently unable to rule out a procedural variance as a contributing factor to the reported event which does not represent a device malfunction.Without the requested clinical information a thorough medical investigation cannot be rendered.According to the report, the surgeon changed 1 shell to a larger size after three different liners would not seat.There were no deviations noted on the dimensional evaluation of the returned shell.Based on the information provided, the surgeon reamed an additional 2mm and a 60mm cup was used to complete the procedure.It was reported there was a delay of > 30-minutes without a significant impact on the patient.Therefore, no further clinical/medical assessment is warranted at this time.Should any additional medical information be provided, this complaint would be re-assessed.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of risk management files and instructions for use found that the reported failure was documented appropriately.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 58MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10649442
MDR Text Key210421650
Report Number1020279-2020-05251
Device Sequence Number1
Product Code MBL
UDI-Device Identifier03596010598271
UDI-Public03596010598271
Combination Product (y/n)N
PMA/PMN Number
K070756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71335558
Device Catalogue Number71335558
Device Lot Number19KW07210
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2021
Date Manufacturer Received09/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
71331758 / 15CT52436; 71332758 / 18MM01522; 71335758 / 17KM00158
Patient Outcome(s) Other; Required Intervention;
-
-