MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VHR
Device Problem Defective Device (2588)
Patient Problem No Patient Involvement (2645)
Event Type  Malfunction  
Manufacturer Narrative

Device was received and evaluated. Evaluation determined that the distal end was observed to be separated from the units bending section. Distal end plastic cover observed to be separated from bending section. The identified parts needs to be replaced. Device was placed for repair. The reported issue was confirmed. The root cause of the failure likely attributed to users handling issue and or maintenance issue.

 
Event Description

It was reported that the user site facility requested and returned the device for a repair to change the units tube/bending section as a scheduled preventive maintenance. There was no patient involvement on this report, no user harm was reported.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCYSTO-NEPHRO VIDEOSCOPE
Type of DeviceCYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10653161
MDR Text Key223662722
Report Number8010047-2020-07431
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeCA
PMA/PMN NumberK062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial,Followup
Report Date 11/24/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/08/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberCYF-VHR
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer06/22/2020
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received10/29/2020
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured02/27/2012
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-