BOSTON SCIENTIFIC CORPORATION INVIVE PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number V173
Device Problems Incorrect, Inadequate or Imprecise Resultor Readings (1535); Telemetry Discrepancy (1629)
Patient Problems No Code Available (3191); Insufficient Information (4580)
Event Date 09/16/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the patient implanted with this cardiac resynchronization therapy pacemaker (crt-p) presented to the emergency department. Difficulty was encountered interrogating the device with a programmer. Upon further troubleshooting, it was found the device had reverted into safety mode. The patient was pacemaker dependent. Subsequently, the patient underwent a revision procedure immediately and the crt-p device was successfully explanted and replaced. The device is not expected to be returned. No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameINVIVE
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key10653785
MDR Text Key210568644
Report Number2124215-2020-20318
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S079
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/18/2014
Device Model NumberV173
Device Catalogue NumberV173
Device Lot Number101974
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/28/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2016-2021

Patient Treatment Data
Date Received: 10/08/2020 Patient Sequence Number: 1
-
-