• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 10622
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/20/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr.: synergy ous mr 4.00 x 32mm stent delivery system catheter was returned for analysis.A visual examination of the stent found damage.Struts lifted and bent from mid-region of stent.The undamaged stent outer diameter was measured and the result is within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 25sep2020.It was reported that crossing difficulties were encountered.The 95% stenosed target lesion was located in the severely tortuous and severely calcified left anterior descending artery.A 4.00 x 32 synergy ii drug-eluting stent was but failed to cross the lesion.The procedure was completed with another of the same device.There were no patient complications nor injuries reported.However, returned device analysis revealed a stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10672162
MDR Text Key211297722
Report Number2134265-2020-14140
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/22/2020
Device Model Number10622
Device Catalogue Number10622
Device Lot Number0023905867
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/25/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
-
-