BLUE BELT TECHNOLOGIES NDI CAMERA POLARIS SPECTRA ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number PFSR200027
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/21/2020
Event Type  malfunction  
Event Description
It was reported that during a navio procedure, the camera start to show orange led, and warning message appeared on the screen, telling that camera has no full functionality, and reduced field of view (infrared). The surgery was completed by using the same device with a minimum delay (less than 30 minutes). This is the fourth of four incidents. No other complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNDI CAMERA POLARIS SPECTRA
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
richard confer
2828 liberty ave
suite 100
pittsburgh, PA 15222
7634524980
MDR Report Key10675796
MDR Text Key211234651
Report Number3010266064-2020-01866
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/21/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberPFSR200027
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/17/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-