MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number ACU0T0
Device Problem Mechanics Altered (2984)
Patient Problem Insufficient Information (4580)
Event Date 09/02/2020
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.(b)(4).
 
Event Description
A health professional reported that during an intraocular lens (iol) implant surgery, the plunger went over the lens.The device was exchanged to complete the case.There was no patient impact.
 
Manufacturer Narrative
The device with the lens was returned.The plunger lock and lens stop have been removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger has advanced to the middle of the lens loading area.The plunger has overrode the lens.The lens was located in the lens loading area.The trailing haptic is folded onto the optic along the left side of the plunger.The leading haptic is ahead of the optic.No damage was observed to the device.All product and batch history records are quality reviewed prior to product release.Viscoelastic was not provided.It is unknown if the qualified product was used.The root cause could not be determined for the reported issue.The used device was evaluated.A plunger override was observed in the lens loading area.It is unknown if a qualified viscoelastic was used.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key10677585
MDR Text Key213174464
Report Number1119421-2020-01460
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2022
Device Model NumberACU0T0
Device Lot Number12765958
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2020
Date Manufacturer Received12/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-