Model Number ELI380LD-L
Device Problem Wireless Communication Problem (3283)
Patient Problem No Patient Involvement (2645)
Event Type  Malfunction  
Manufacturer Narrative

The eli 380 device is indicated for use to acquire, analyze, display, and print electrocardiograms. Device is indicated for use to provide interpretation of the data for consideration by a physician. Device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis. The interpretations of ecg offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data. Device is indicated for use on adult and pediatric populations. The device is not intended to be used as a vital signs physiological monitor. If the eli 380 device loses connection to the wireless network, the device is unable to transmit or receive ekg¿s. This failed connection may result in a delay of patient treatment, which may cause or contribute to a serious injury or death. Therefore, hillrom is reporting this alleged connection failure as a product malfunction. The hillrom field service engineer installed the new software on the eli380 device to upgrade the device to a new software version to correct the reported issue. Based on this information, no further action is required at this time.

 
Event Description

Hillrom received a report from the account stating that the eli 380 device would drop connection to the wireless network. The device was located at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as complaint #: (b)(4).

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLNR, ELI 380 TOUCHSCREEN DISPLAY, LAIRD
Type of DeviceELECTROCARDIOGRAPH
Manufacturer (Section D)
MORTARA INSTRUMENT
7865 n 86th street
milwaukee WI 53224
Manufacturer Contact
emily mitchell
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key10678368
MDR Text Key211332153
Report Number2183461-2020-00021
Device Sequence Number1
Product Code DPS
Combination Product (Y/N)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,USER F
Reporter Occupation
Type of Report Initial
Report Date 09/16/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/14/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator NO INFORMATION
Device MODEL NumberELI380LD-L
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/16/2020
Was Device Evaluated By Manufacturer? Yes
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-