OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problems Erratic or Intermittent Display (1182); Output Problem (3005)
Patient Problem No Patient Involvement (2645)
Event Type  Malfunction  
Manufacturer Narrative

The device was returned to olympus for evaluation and during testing it was found that the device had intermittent video caused by a defective video port. During inspection of the device it was found that the contact pins are worn out inside the video connection. Olympus performed service and replaced the video connector and the device passed all functional testing. Olympus will continue to monitor complaints for this device.

 
Event Description

The user facility reported that the cv-180 has a intermittent video. There was no patient involvement associated to this report.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameEVIS EXERA II VIDEO SYSTEM CENTER
Type of DeviceVIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10680731
MDR Text Key223663350
Report Number8010047-2020-07690
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,USER F
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/02/2021
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/14/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator OTHER
Device MODEL NumberCV-180
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer09/28/2020
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received12/04/2020
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured07/24/2016
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-