• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-82
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
There was no known patient involvement.The heater-cooler 16-02-82 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Through follow-up communication livanova learned that the device was cleaned as per the instruction for use and that it placed inside the operating theatre during use with the fan opposite to the patient and at an estimated distance of 3-4 meters from the surgery field.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t device was found to be contaminated with an unknown type of mycobacterium.There is no known patient involvement.
 
Manufacturer Narrative
H10: the hospital microbiology chief had requested to proceed to the genotyping of the ntm.These results are not yet available.Through follow up communication livanova learned that following the reported contamination result, the hospital performed three (3) more disinfection cycle in a row followed by another water sampling.The results are pending to livanova.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: laboratory test results were provided to livanova confirming that the device was contaminated with mycobacterium chimaera.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key10682283
MDR Text Key211442775
Report Number9611109-2020-00573
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-82
Was the Report Sent to FDA? No
Date Manufacturer Received08/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
-
-