Model Number V60 PLUS
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 15oct2020.
 
Event Description
The customer reported that the unit turns off.The customer contacted product support and requested for service repair.The customer reported that the unit was in use on a patient, but there was no patient harm reported.
 
Manufacturer Narrative
G4: 12oct2020.B4: 20oct2020.Per field service engineer (fse) the customer reported problem was verified.After several tests it was not possible to reproduce the problem.Performed operational, functional and safety tests.The equipment complies with factory specifications.No part was replaced.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key10687342
MDR Text Key213467249
Report Number2031642-2020-03736
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838089280
UDI-Public(01)00884838089280
Combination Product (y/n)N
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 09/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60 PLUS
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER
-
-