• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-82
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/17/2020
Event Type  malfunction  
Manufacturer Narrative
There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Through follow-up communication livanova learned that the device was cleaned as per the instruction for use and that it was placed inside the operating theatre during use with the fan opposite to the patient at an estimated distance of four (4) meters from the surgery field.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t device was found to have a bacterial count higher than the prescribed limits.The laboratory report confirming the reported contamination has been provided to livanova.There is no known patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10688677
MDR Text Key211648143
Report Number9611109-2020-00581
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-82
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/17/2020
Date Device Manufactured04/24/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
-
-