TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
|
Back to Search Results |
|
Model Number 4415 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/23/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during the use of the device for a non-clinical activity, there were significant changes in temperature and the unit took too long to heat.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.Per the manufacturer's subsidiary, the distributor's service engineer completed the preventive maintenance steps and ensured that the heater cooler operated to the manufacturer's specification.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
Per subsidiary, after following the diagnostic steps provided by the manufacturer the service engineer was able to get the thermostat of the heater cooler working again.
|
|
Search Alerts/Recalls
|
|
|