Model Number 4415
Device Problem Temperature Problem (3022)
Patient Problem No Patient Involvement (2645)
Event Date 09/23/2020
Event Type  malfunction  
Event Description
It was reported that during the use of the device for a non-clinical activity, there were significant changes in temperature and the unit took too long to heat.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.Per the manufacturer's subsidiary, the distributor's service engineer completed the preventive maintenance steps and ensured that the heater cooler operated to the manufacturer's specification.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
Per subsidiary, after following the diagnostic steps provided by the manufacturer the service engineer was able to get the thermostat of the heater cooler working again.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key10690929
MDR Text Key211906634
Report Number1828100-2020-00394
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-