Catalog Number NPFS02000
Device Problem Application Program Problem: Parameter Calculation Error (1449)
Patient Problems Injury (2348); Deformity/ Disfigurement (2360)
Event Date 09/23/2020
Event Type  Injury  
Event Description
It was reported that, during a navio-assisted ukr procedure, the lug holes were not completed as some blue areas remained, resulting an incomplete seating of the component and, therefore, pushing into slight valgus. It caused incorrect final alignment. The problem was resolved by removing more tibia. Surgery was not delayed beyond 30 min.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNAVIO SURGICAL SYSTEM US
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
suite 100
austin, TX 78735
5123913905
MDR Report Key10690957
MDR Text Key211696016
Report Number3010266064-2020-01878
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberNPFS02000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/07/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 10/16/2020 Patient Sequence Number: 1
-
-