Model Number 22042
Device Problems Premature Discharge of Battery (1057); Incorrect Measurement (1383)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/09/2020
Event Type  malfunction  
Manufacturer Narrative
The product has been returned to masimo for evaluation.When the investigation is complete a follow up report will be submitted.
 
Event Description
The customer reported the device is not holding a charge or providing good readings.No patient impact or consequences were reported.
 
Event Description
The customer reported the device is not holding a charge or providing good readings.No patient impact or consequences were reported.
 
Manufacturer Narrative
Additional manufacturing narrative: other, other text: the returned device was evaluated.External visual inspection showed no damage.The device was unable to power on using battery power.The unit powered on and off correctly using ac power and the device turns off immediately if ac power is removed.When powered on the unit was able to obtain readings and alarmed audibly and visually under alarm conditions.Spo2 and pulse rate simulator tests were performed and no issues related to measurement accuracy were identified.Internal inspection found the housing screw mount had broken off.The battery charging circuit output was measured low when connected to ac power.A known good battery was installed and the device was able to power on and off using battery power.The customer's battery was returned to the device and was now able to charge.The customer's complaint regarding readings was not duplicated however the complaint related to not holding a charge was duplicated.A service history record review reveals that this unit was in the field for over three (3) years with no previous reported issues related to this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-8, HORIZONTAL
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 52 DISCOVERY
52 discovery
irvine CA 92618
MDR Report Key10692131
MDR Text Key211733713
Report Number3011353843-2020-00168
Device Sequence Number1
Product Code DQA
UDI-Device Identifier00843997005647
UDI-Public00843997005647
Combination Product (y/n)N
PMA/PMN Number
K053269
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number22042
Device Catalogue Number9190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2020
Was the Report Sent to FDA? No
Date Manufacturer Received10/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-