Model Number M0063903100 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/21/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The complainant was unable to provide the suspect device lot number; therefore, the expiration and device manufacture dates are unknown.However, the complainant reported that the device was not expired.(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a stone cone nitinol retrieval coil was used in the ureter during a ureteroscopy (urs) procedure performed on (b)(6) 2020.According to the complainant, during procedure, the stone cone was inserted into the ureter and was opened behind a stone.However, when closed and pulled back, the tip was found to be detached into the uterus.The detached tip was retrieved with a forceps.Lithotripsy was performed with a non-bsc thulium laser device, and the procedure was completed.There were no patient complications as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
|