These machines had system failures and required repair due to an arps system failure alarm (b1273).
This error does not allow any actions other than discontinue treatment.
(b)(4).
Additionally, there were two machines that had the auto-effluent cassette loaded, but it would not release for removal during shutdown.
One occurred during class and (baxter educator) was able to help resolve it through restarting the machine and re-shutting it down.
The second one could not be resolved.
We ended up working it off of the side.
The machine was able to be set up on a subsequent use without difficulty.
Neither of these events created a work order but were witnessed during class time by the educator.
While baxter engineering was in-house, he was asked to replace/repair the rubber stopper on two of the thermax blood warmers.
The rubber stopper at the bag insertion area becomes loose and prevents the blood warmer bag from advancing into the warmer.
He was able to secure both of these stoppers while present for the other repairs.
Manufacturer response for baxter prismax crrt machine, (brand not provided) (per site reporter).
|