Model Number 45007 |
Device Problem
Entrapment of Device (1212)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/30/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that the catheter became stuck on the wire.A 2.4 mm jetstream xc atherectomy catheter was selected for a procedure in the left superficial femoral artery (sfa).During the procedure, the jetstream catheter was advanced over a thruway guidewire and the jetstream ran for 7:23 minutes.Then, catheter became stuck on the guidewire and it could not be removed.The catheter and the wire were removed together from the patient.There were no patient complications reported.
|
|
Event Description
|
It was reported that the catheter became stuck on the wire.A 2.4 mm jetstream xc atherectomy catheter was selected for a procedure in the left superficial femoral artery (sfa).During the procedure, the jetstream catheter was advanced over a thruway guidewire and the jetstream ran for 7:23 minutes.Then, catheter became stuck on the guidewire and it could not be removed.The catheter and the wire were removed together from the patient.There were no patient complications reported.
|
|
Manufacturer Narrative
|
Corrections to field h6: method, results, and conclusion.Device analysis by mfr: the returned product consisted of a jetstream xc 2.4.The guidewire was in the device when returned.The device and the catheter shaft were analyzed for damage.The baton and the electrical connections were cut off from the customer site.The catheter shaft showed 1 kink and multiple areas of buckling of the shaft.The kink was located 63.5cm from the tip.The guidewire that was in the device was not sticking out of the distal end.The guidewire was sticking out of the proximal end approximately 220cm.The wire was pulled from the device with difficulty and showed tip damage when removed.It looked like the jetstream device was run of the coils of the tip.The device could not be functionally tested due to the baton and electrical connections being cut from the device.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
|
Manufacturer Narrative
|
Device analysis by mfr: the returned product consisted of a jetstream xc 2.4.The guidewire was in the device when returned.The device and the catheter shaft were analyzed for damage.The baton and the electrical connections were cut off from the customer site.The catheter shaft showed 1 kink and multiple areas of buckling of the shaft.The kink was located 63.5cm from the tip.The guidewire that was in the device was not sticking out of the distal end.The guidewire was sticking out of the proximal end approximately 220cm.The wire was pulled from the device with difficulty and showed tip damage when removed.It looked like the jetstream device was run of the coils of the tip.The device could not be functionally tested due to the baton and electrical connections being cut from the device.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
|
Event Description
|
It was reported that the catheter became stuck on the wire.A 2.4 mm jetstream xc atherectomy catheter was selected for a procedure in the left superficial femoral artery (sfa).During the procedure, the jetstream catheter was advanced over a thruway guidewire and the jetstream ran for 7:23 minutes.Then, catheter became stuck on the guidewire and it could not be removed.The catheter and the wire were removed together from the patient.There were no patient complications reported.
|
|
Search Alerts/Recalls
|