Catalog Number UNK HIP ACETABULAR CUP
Device Problems Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Patient Problem No Code Available (3191)
Event Date 10/01/2020
Event Type  Injury  
Manufacturer Narrative

Product complaint # (b)(4). Investigation summary: no device associated with this report was received for examination. A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided. The information received will be retained for trend analysis, post market surveillance, or other events within the quality system. (b)(4).

 
Event Description

The patient was revised due to liner disassociation from cup. Cup had wear from head on edge. Cup was deformed so replaced it with competitor's mom cup. No surgical delay. Doi: (b)(6) 2015; dor: (b)(6) 2020; right hip.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN HIP ACETABULAR CUP
Type of DeviceHIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic drive
warsaw, IN 46581-0988
6107428552
MDR Report Key10710809
MDR Text Key212283039
Report Number1818910-2020-22834
Device Sequence Number1
Product Code KWA
Combination Product (Y/N)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial
Report Date 10/01/2020
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received10/21/2020
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue NumberUNK HIP ACETABULAR CUP
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/01/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 10/21/2020 Patient Sequence Number: 1
Treatment
12/14 ARTICUL 40MM M SPEC+1.5; UNK HIP ACETABULAR LINER POLY; UNK HIP BONE SCREW
-
-