MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number RONYX22522UX
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Myocardial Infarction (1969); Occlusion (1984); Injury (2348); Obstruction/Occlusion (2422); Insufficient Information (4580)
Event Date 10/12/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
A resolute onyx rx des was attempted to be used to treat a mildly tortuous, mildly calcified lesion exhibiting 90% stenosis in the 1st obtuse marginal.The proximal cx and the distal left main were also stented.There was no damage noted to the packaging.There was no damage noted when removing the device from the hoop.The device was inspected with no issues noted.Negative prep was not performed.The lesion was pre dilated.The device did not pass through a previously deployed stent.Excessive force was not used during delivery.It was reported that the stent dislodged during delivery to the lesion and flow was lost in the lad.The dislodged stent was crushed in the lad, in order to save the lad.It was also reported that mi and vessel occlusion occurred due to the device.A 2.75 x 18 mm stent was then implanted in the proximal lad and a 3.0 x 8 mm stent in the left main/ostial lad.Patient was sent to another hospital ((b)(6) hospital) and required bypass surgery two days later.The patient is reported to be alive with injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: it was later reported that the mi was treated with an intra-aortic balloon pump (iabp) insertion for hemodynamic stability.The patient was also transferred to (b)(6) medical center for further care following a cabg procedure.The patient was discharged to home from (b)(6).The patient was transferred to (b)(6) hospital for a cabg procedure ((b)(6)2020).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key10711157
MDR Text Key212287754
Report Number9612164-2020-04028
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier00643169556737
UDI-Public00643169556737
Combination Product (y/n)N
PMA/PMN Number
P160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 12/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/24/2021
Device Model NumberRONYX22522UX
Device Catalogue NumberRONYX22522UX
Device Lot Number0009983488
Was Device Available for Evaluation? No
Date Manufacturer Received12/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age59 YR
Patient Weight129
-
-