Model Number 2010-01-040
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Code Available (3191)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Event Description
It was reported that the kincise unit broke the attachment off the broach while extracting the broach.There was a surgical delay of 30 minutes.
 
Manufacturer Narrative
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Patient code: no code available (3191) is used to capture insufficient information.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS BROACH SZ 4
Type of Device
BROACHES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10714452
MDR Text Key212482098
Report Number1818910-2020-22995
Device Sequence Number1
Product Code KWL
UDI-Device Identifier10603295393665
UDI-Public10603295393665
Combination Product (y/n)N
PMA/PMN Number
K150862
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2010-01-040
Device Catalogue Number201001040
Device Lot NumberPG274010
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/08/2020
Date Manufacturer Received10/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-