• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Model Number 283512
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/08/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint#: (b)(4).To date, the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.Udi: (b)(4).
 
Event Description
It was reported by the sales rep via phone that before a rotator cuff repair procedure the o-ring of the fms tornado micro handpiece with buttons popped out of the shaver.Another devices was used to complete the procedure.No patient consequences, or surgical delay reported.The device is available to be returned for evaluation.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary :he device was received and evaluated at the service center.The reported complaint that the o-ring of the fms tornado micro handpiece with buttons popped out of the shaver was confirmed.It was found that the device was physically damaged.Further, the motor, motor cable and the o-rings were found to be defective.There were also minor scratches on the device identified during decontamination.The defective parts were replaced and the device was tested and found to be working according to specifications.The physical damage to the device and the minor scratches are most likely a result of user mishandling of the device.However, given the information provided, we cannot discern a definitive root cause of the other identified failures.A manufacturing record evaluation was performed for the finished device [serial number : (b)(6), non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10715673
MDR Text Key212475064
Report Number1221934-2020-03146
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 10/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/12/2020
Date Manufacturer Received12/04/2020
Removal/Correction NumberN/A
Patient Sequence Number1
-
-