Catalog Number PMKEM1
Device Problem Mechanical Problem (1384)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction has caused file separation. Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur. As such, this event meets the criteria for reportability per 21 cfr part 803. Console tested the torque using calibrated torque test fixture; tested okay. Foot pedal came in shorted; replaced the switch activation ball and aligned; tested okay. Motor m250t-27366, and the pwr cord both tested okay.
 
Event Description
In this event it was reported that a promark motor was not auto-reversing properly (too quickly). The event outcome is unknown as of this mdr evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROMARK ENDODONTIC MOTOR
Type of DeviceHANDPIECE, DIRECT DRIVE, AC-POWERED
Manufacturer (Section D)
TULSA DENTAL PRODUCTS LLC
608 rolling hills drive
johnson city TN 37604
Manufacturer (Section G)
ASEPTICO, INC
8333 216th st se
woodinville WA 98072
Manufacturer Contact
karl nittinger
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494424
MDR Report Key10715839
MDR Text Key213900784
Report Number2320721-2020-00147
Device Sequence Number1
Product Code EKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111078
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 03/31/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberPMKEM1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/21/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-