• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ECLIPSE FILTER SYSTEM - FEMORAL; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ECLIPSE FILTER SYSTEM - FEMORAL; VENA CAVA FILTER Back to Search Results
Catalog Number EC500F
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Abdominal Pain (1685)
Event Date 05/22/2019
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation; however, medical records were provided for review.The investigation of the reported event is currently underway.Medical device expiry date: 01/2014.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient in conjunction with or before orthopedic procedure.At some time post filter deployment, it was alleged that the filter perforated.The device has not been removed and there were no reported attempts made to retrieve the filter.The patient reportedly experienced abdominal pain; however, the current status of the patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: a lot history review was performed.This is the only complaint to date for this lot number.Therefore, a device history record review is not required.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Approximately eight years and one month later, computed tomography (ct) revealed anchoring struts extended beyond the vessel wall distally.There was no tilt, bend, or breakage of the device.Therefore, the investigation is confirmed for perforation of the inferior vena cava (ivc).Based on the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H10:d4 (expiry date: 01/2014), g4.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient in conjunction with or before orthopedic procedure.At some time post filter deployment, it was alleged that the filter perforated.The device has not been removed and there were no reported attempts made to retrieve the filter.The patient reportedly experienced abdominal pain; however, the current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECLIPSE FILTER SYSTEM - FEMORAL
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key10725392
MDR Text Key212687582
Report Number2020394-2020-06019
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K093659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 11/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEC500F
Device Lot NumberGFUL1901
Was Device Available for Evaluation? No
Date Manufacturer Received10/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age72 YR
-
-