LIVANOVA DEUTSCHLAND ELECTRICAL VENOUS OCCLUDER (EVO); CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 12-80-10
Device Problem Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/24/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the electrical venous occluder (evo).The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The affected part was returned to livanova (b)(4) for a detailed investigation.The investigator could reproduce the reported issue.The problem could be traced to faulty mechanical tolerances at the bearing seat of the linear actuator.The parts will be replaced to solve the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a electrical venous occluder (evo) was giving an error code associated to the encoder during procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRICAL VENOUS OCCLUDER (EVO)
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10725947
MDR Text Key212721183
Report Number9611109-2020-00592
Device Sequence Number1
Product Code DXC
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K082344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 10/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number12-80-10
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/24/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/11/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-