Model Number 25-40-45
Device Problem Gas Output Problem (1266)
Patient Problem No Patient Involvement (2645)
Event Date 09/24/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement. Livanova (b)(4) manufactures the s5 gas blender system. The incident occurred in (b)(6). The affected device was returned to livanova (b)(4) for a detailed investigation. The reported issue could be reproduced. The problem could be traced to an electronic defective component. The front plate where the faulty component was installed will be replaced to solve the problem.
 
Event Description
During a maintenance activity a livanova field service engineer experienced an error code associated to discrepancies of set and actual value of gas output on a s5 gas blender system. The failure occurred only once. There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameS5 GAS BLENDER SYSTEM
Type of DeviceGAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10725956
MDR Text Key214553184
Report Number9611109-2020-00593
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/21/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number25-40-45
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received01/05/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

-
-