DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Failure to Charge (1085); Connection Problem (2900)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge service territory manager (stm) was dispatched to evaluate the reported issue and observed that pin on ac power plug was loose.The stm replaced the retractable cord assembly.In addition batteries were verified and were charging properly.Subsequently the stm performed all calibration, functional, and safety tests, the unit passed all per factory specifications.The iabp was then returned to customer and cleared for clinical use.
|
|
Event Description
|
It was reported that during routine check, the cardiosave intra-aortic balloon pump (iabp) was not charging and the plug seemed damaged.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during routine check, the cardiosave intra-aortic balloon pump (iabp) was not charging and the plug seemed damaged.There was no patient involvement, and no adverse event reported.
|
|