Model Number 319.006
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date, the depth gauge was found broken in sterile processing while cleaning.There was no patient involvement.This report is for one.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6- the depth gauge for 2.0mm and 2.4mm screws (part #: 319.006, lot #: 9928856) was received at us cq.Upon visual inspection, it was observed that the needle component had broken off of the center shaft of the device.The broken off needle was not received at us cq.The protection sleeve was not returned.This complaint was confirmed as the needle component had broken off of the center shaft of the device.While no definitive root cause could be determined, it is possible the device encountered unintended forces during use.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot part # 319.006, synthes lot # 9928856, release to warehouse date; 11 nov 2015, manufactured by synthes jennersville, no ncr's were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10727063
MDR Text Key212747402
Report Number2939274-2020-04802
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier10886982189943
UDI-Public10886982189943
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number319.006
Device Catalogue Number319.006
Device Lot Number9928856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2020
Date Manufacturer Received11/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-