A synergy ous mr 3.00 x 38 mm stent delivery system was returned for analysis.A visual examination of the stent found that the stent was damaged from mid-section of stent to the distal end, with stent struts lifted from crimped stent positions and misaligned.The undamaged crimped stent outer diameter measured and the result was within maximum crimped stent profile measurement.Stent damage most likely occurred when the stent struts got caught in the calcification.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the tip showed signs of damage on the distal edges of the tip.Tip damage most likely occurred when the tip was pushed against a restriction.A visual and tactile examination of the hypotube found multiple kinks along several locations of the hypotube shaft.This type of damage is consistent with excessive force that could have been applied to the delivery system.A visual and tactile examination of the outer extrusion, mid-shaft section, and visual examination of the inner extrusion found no issue with the extrusion shaft.No other issues were identified during the product analysis.
|
Reportable based on device analysis completed on 03oct2020.It was reported that crossing difficulties were encountered.The 80% stenosed target lesion was located in the moderately tortuous and severely calcified right coronary artery.After the lesion was pre-dilated, a 3.00 x 38 synergy ii drug-eluting stent was advanced, but failed to cross the lesion.The procedure was completed with another of the same device.There were no patient complications reported.However, returned device analysis revealed stent damage.
|