Since the subject device has not been returned to olympus medical systems corp.
(omsc) for the evaluation, the exact cause of the reported event could not be conclusively determined at this time.
If additional information becomes available, this report will be supplemented.
|
Olympus medical systems corp.
(omsc) was informed from the user that the image of the subject device was not displayed and the display color was pink at the preparation for use.
At the incoming inspection for the repair, olympus india checked the subject device.
(b)(4) found that the subject device could not be started up and all leds on the front panel were blinking continuously due to the printed circuit board failure of the subject device, and the fans of the power supply unit were not worked due to the power supply unit failure of the subject device.
There was no report of patient injury associated with this event.
|