BAXTER HEALTHCARE CORPORATION EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 7N8300 |
Device Problem
Component Missing (2306)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/17/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the luer of an one-link non-dehp microbore catheter extension set was missing.This was identified when the package was opened, prior to patient use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information was added to h3, h4 and h6.H10: the actual device was received for evaluation.Visual inspection was performed to the actual sample using the naked eye which revealed that the next - gen luer activated device was missing in the set.Functional testing including clear passage and pressure testing were performed and no issues were noted.The reported condition of missing component was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|