The lot was manufactured from july 27, 2020 - july 28, 2020.Eleven (11) companion samples were received for evaluation.Unaided visual inspection was performed did not identify any abnormalities that could have contributed to the reported condition.All of the clamps were damage when an excessive force is required to close the clamps.During analysis, two loose tabs broke off inside the closed damaged clamp of 5 samples and 1 loose tab broken off inside the closed damaged clamp of 6 samples.The reported condition was verified.The cause of the condition was due to a supplier manufacturing issue.This issue is being further investigated.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|