Device evaluated by mfr.: promus element plus, mr, ous 4.00x38 mm stent delivery system was returned for analysis.A visual examination of the stent found stent struts on proximal end lifted and pulled in a distal direction.The stent outer diameter was measured using snap gauge and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed signs of damage.A visual and tactile examination of the hypotube found no issues.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|