BOSTON SCIENTIFIC CORPORATION PROMUS PREMIER SELECT; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 10663
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/04/2020
Event Type  malfunction  
Manufacturer Narrative
A visual examination of the stent found signs of stent damage.Stent struts from the proximal region of the stent were noted to be lifted and pulled in a distal direction.The undamaged stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed signs of distal tip damage.A visual and tactile examination of the hypotube found no issues.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 14-oct-2020.It was reported that crossing difficulties were encountered.The 16 x 3.00mm target lesion was located in the moderately tortuous left anterior descending artery.A 16 x 3.00mm promus premier select drug-eluting stent was advanced but failed to cross the lesion.The device was removed from the patient's body and the procedure was completed with another of same device.No patient complications reported and the patient's status was stable.However, returned device analysis revealed stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER SELECT
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10742114
MDR Text Key213260611
Report Number2134265-2020-14349
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeTU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/29/2022
Device Model Number10663
Device Catalogue Number10663
Device Lot Number0025160981
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/14/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-