ABBOTT VASCULAR XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1070350-12
Device Problem Material Separation (1562)
Patient Problem Vessel Or Plaque, Device Embedded In (1204)
Event Date 09/29/2020
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history identified no similar incidents from this lot. The investigation was unable to determine a conclusive cause for the reported material separation/stent break; however, the subsequent treatments appear to be related to circumstances of the procedure. There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event Description
It was reported that the procedure was to treat a moderately tortuous, heavily calcified right coronary artery (rca). During an implantation of a 3. 5x38mm xience xpedition stent, the struts of the stent became separated. The stent was crushed into the vessel and another xience xpedition was used to successfully complete the procedure. There was no clinically significant delay. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameXIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of DeviceDRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10742121
MDR Text Key213382259
Report Number2024168-2020-08936
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/03/2022
Device Catalogue Number1070350-12
Device Lot Number9022741
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/05/2020
Is This a Reprocessed and Reused Single-Use Device? No

Patient Treatment Data
Date Received: 10/27/2020 Patient Sequence Number: 1
-
-