Catalog Number 08429324160
Device Problems Low Test Results (2458); Non Reproducible Results (4029)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/01/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.
 
Event Description
The initial reporter received questionable tsh results on a cobas 6000 e601 module serial number, (b)(4).The initial result was 0.028 iu/ml and the repeated result was 2.33 iu/ml.The initial result was reported outside of the laboratory.The repeated result was believed to be correct.
 
Manufacturer Narrative
The field service engineer found damaged measuring cells that he replaced.The customer calibrated and ran controls.Calibration and qc data provided by the customer were acceptable.The alarm trace from the analyzer contained an abnormal aspiration alarm, indicating possible poor sample quality.The investigation determined that the issue found by the fse was the root cause of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TSH ELECSYS COBAS E 200 V2
Type of Device
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key10742457
MDR Text Key214803346
Report Number1823260-2020-02676
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 11/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Catalogue Number08429324160
Device Lot Number466153
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age57 YR
-
-