BOSTON SCIENTIFIC CORPORATION VISIONIST X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U228 |
Device Problems
Failure to Capture (1081); Over-Sensing (1438); Pacing Problem (1439); High Capture Threshold (3266)
|
Patient Problem
Fall (1848)
|
Event Date 07/24/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Investigation of the available information determined this device exhibited oversensing of noise generated by the minute ventilation (mv)/ respiratory sensor that is related to a high impedance condition.Please see the description for more information regarding the specific circumstances of this event.This device was included in the recent minute ventilation sensor signal oversensing advisory population.
|
|
Event Description
|
It was reported that this patient experienced pacing inhibition resulting in asystole and collapsed at home.The patient was subsequently hospitalized and boston scientific technical services were contacted.It was discussed that a signal artifact monitoring (sam) episode had occurred due to oversensing of the minute ventilation (mv) feature signal from a competitor's lead.It was recommended to turn this mv feature off and perform provocative maneuvers to exclude lead impairment.At this time, the system remains implanted and in service.The patient was stable with no additional consequences reported.
|
|