A steris service technician arrived onsite to inspect the table and found the table's floor lock feet were too close to the floor resulting in the table not being able to re-lock.The technician adjusted the table's feet to the correct height, tested the unit, confirmed it was operating according to specifications, and returned it to service.The reported event is attributed to user error as facility personnel should not have unlocked the 3085 sp surgical table while a patient was on the table.The 3085 sp surgical table operator manual states (1-1), "warning - tipping hazard: do not place patient on the table unless floor locks are engaged.Do not release floor locks while patient is on table." the technician counseled user facility personnel including the doctor on the proper use and operation of their 3085 sp surgical table, specifically to not release the floor locks while a patient is on the table.No additional issues have been reported.
|