Model Number 52969
Device Problem Failure to Cut (2587)
Patient Problem Insufficient Information (4580)
Event Date 10/15/2020
Event Type  malfunction  
Manufacturer Narrative
Coopersurgical, inc.Is currently investigating the reported conditon.
 
Event Description
Ref: e-complaint: (b)(4).Dr.Stated unit cuts out while cutting.Order: (b)(4).1216677-2020-00250 leep system 1000 esu gen 52969 e-complaint: (b)(4).
 
Manufacturer Narrative
Investigation: x-inspect returned samples.*analysis and findings: complaint # (b)(4).Distribution history: this complaint unit was manufactured at csi on 03/17/2004 under wo # (b)(4) and shipped on 5/10/2004.Manufacturing record review: a review of the device history record could not be performed as the record could not be located at this time.However, at the time of manufacture, records from each unit are reviewed to ensure that product meet all specifications.Should the device history record be located, this complaint will be amended accordingly.Incoming inspection review: n/a.Service history record: no additional service history records found for this unit.Historical complaint review: a review of the 2-year complaint history showed similar reported complaint conditions.Product receipt: the complaint unit was returned on a repair log 95099.Visual evaluation: visual examination of the complaint unit revealed no physical damage.Functional evaluation: unit was not functioning properly.Root cause: the diaphragm is the mechanism to make contact (for electrical current) to supply power to the unit.Air displacement pushes on a piston via the diaphragm into a switch to turn the power on.The foot pedal creates the air displacement to the pneumatic switch mechanism located in the housing of the unit.The diaphragm breaks down and loses its seal and cannot function properly.The diaphragm is described as a rubber, possibly a latex.Given the appearance of a bad diaphragm it appears to have been exposed to excessive stimuli.Materials like latex are flexible hence the use in this application, but its' elastic properties can alter over time as well.The root cause for this complaint condition is component related to the diaphragm.*correction and/or corrective action / *preventative action activity coopersurgical service and repair replaced the diaphragm on the unit, tested it and returned it to the customer.Engineering has successfully tested a replacement material made of silicone for use in repairs going forward, eng-test-10341-r.The ifu was also updated to add a safety check via (b)(4).A service bulletin was issued to existing customers informing them to check for this issue and return the unit if needed.All product in fg and sk, as applicable, were reworked to replace the previous versions of the dfus on all applicable products.The repaired unit will have been repaired with this new silicone material.No further training required at this time.
 
Event Description
Ref : (b)(4).Dr.Stated unit cuts out while cutting.Order: (b)(4).1216677-2020-00250 leep system 1000 esu gen 52969 (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEEP SYSTEM 1000 ESU GEN
Type of Device
LEEP SYSTEM 1000 ESU GEN
Manufacturer (Section D)
COOPERSURGICAL, INC.
95 corporate drive
trumbull CT 06611
Manufacturer (Section G)
COOPERSURGICAL, INC.
75 corporate drive
trumbull CT 06611
Manufacturer Contact
michael marone
50 corporate drive
trumbull, CT 06611
4752651665
MDR Report Key10755668
MDR Text Key227973100
Report Number1216677-2020-00250
Device Sequence Number1
Product Code HGI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K952483
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number52969
Device Catalogue Number52969
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/22/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-