Model Number 6534 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/05/2020 |
Event Type
Injury
|
Event Description
|
It was reported that shaft break occurred.The 95% stenosed target lesion was located in the mildly tortuous and severely calcified mid left anterior descending artery.A 30mm x 2.50mm nc quantum apex balloon catheter was advanced for dilation.During removal, the distal portion remained in patient due to a bend.The device was completely removed by trapping the wire and inflating another balloon.The procedure was completed with another of the same device and no further complications were reported.
|
|
Event Description
|
It was reported that shaft break occurred.The 95% stenosed target lesion was located in the mildly tortuous and severely calcified mid left anterior descending artery.A 30mm x 2.50mm nc quantum apex balloon catheter was advanced for dilation.During removal, the distal portion remained in patient due to a bend.The device was completely removed by trapping the wire and inflating another balloon.The procedure was completed with another of the same device and no further complications were reported.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: the device was returned for analysis.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 56.8cm distal of the strain relief.The was blood in the guidewire lumen.The balloon was tightly folded.Microscopic inspection revealed tip damage.Inspection of the remainder of the device presented no other damage or irregularities.
|
|
Search Alerts/Recalls
|