Model Number CV-180
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2020
Event Type  Malfunction  
Manufacturer Narrative

The device has not been returned to olympus medical systems corp. (omsc) for evaluation. In the evaluation of the service department of olympus (thailand) co. , ltd. (oth) the following was confirmed; the reported phenomenon was reproduced. There was a problem with the conversion adapter for outputting the video signal from the device. The device is now outputting all output video signals properly. The exact cause of the reported event could not be conclusively determined at this time. If additional information becomes available, this report will be supplemented.

 
Event Description

During preparation for use before laparoscopic surgery, the device's the sdi port didn't work and the sdi video signal was not sent to the monitor. The endoscopic image was not displayed. The user switched to another video signal port and completed the procedure using the device. The device was used with the olympus light source clv-180. There was no report of patient injury associated with this event.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameEVIS EXERA II VIDEO SYSTEM CENTER
Type of DeviceVIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10761419
MDR Text Key224665864
Report Number8010047-2020-08327
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeTH
PMA/PMN NumberK133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type FOREIGN,USER FACILITY
Reporter Occupation
Type of Report Initial,Followup
Report Date 12/21/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/30/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberCV-180
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received11/27/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured10/17/2011
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-