EDWARDS LIFESCIENCES EDWARDS ESHEATH INTRODUCER SET; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Embolism (1829)
Event Date 09/10/2020
Event Type  Injury  
Manufacturer Narrative
Udi: unknown.The investigation is ongoing.
 
Event Description
As reported via a call to edward lifesciences,  8-9 days post right tf tavr the patient developed a rash on the right foot.An echo and doppler study on both legs performed showed normal results.A dermatologist performed a right foot punch biopsy of skin measuring 2x2x2mm. per the pathology report stated ¿there is basophilic material within a muscular blood vessel.¿ ¿this suggests an embolic phenomenon.¿ the patient was prescribed medication with instructions to stop plavix and the rash subsided. from last communication the patient is doing much better with no more rashes.
 
Manufacturer Narrative
Udi: unknown.  per the instructions for use (ifu), potential risks associated with the overall procedure include thrombus formation, plaque dislodgment, and embolization that may result in myocardial infarction, stroke, distal peripheral occlusion, and/or death.It is the natural tendency of the body to form clot on foreign objects in the vascular space.These patients are anticoagulated for the procedure and interventional best practices mandate meticulous wiping and flushing of the devices to prevent and/or remove clot.The thv training manuals and ifu instruct the operator to administer heparin and maintain the act at = 250 sec.In this case, there was no allegation or indication a device malfunction contributed to this adverse event.The exact cause for the ¿embolic phenomenon¿ could not be determined, however, procedural factors listed above and/or patient factors not provided may have contributed to the event. the ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS ESHEATH INTRODUCER SET
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key10765375
MDR Text Key214152359
Report Number2015691-2020-14194
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received12/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-